8 Technical Data
8.1 Product Parameters
OXYGEN
Oxygen
20.9 - 96%
Measurement
Range: (from a
concentrator)
Oxygen
±1.5 % of full scale at constant temperature
Measurement
and optimal flow*
Accuracy:
Oxygen
0.1% Oxygen
Measurement
Resolution:
See Effects of Low Flow, page 11 of this operating manual
for information on the effects of low flow.
FLOW
Flow Measurement
0 - 10 LPM
Range:
Flow Measurement
±0.2 LPM
Accuracy:
Flow Measurement
0.1 LPM
Resolution:
PRESSURE
PSI
Pressure
0.5 - 50
Measurement Range:
Pressure
±0.5%
Measurement
Accuracy:
1176119-A
KPA
3.4 - 344
±0.5%
Pressure
0.1
Measurement
Resolution:
Response Time:
£ 17 seconds
Warm-up Time:
< 1 second
Operating
59° F - 104° F (15° C - 40° C)
Temperature:
Storage
5° F - 140° F (-15° C - 60° C)
Temperature:
Pressure:
800 - 1000 mBars
Humidity:
0 - 95% (non-condensing)
Power
2-AA Alkaline batteries (2 x 1.5 volts)
Requirements:
³1,100 hours (16,500 read cycles)
Battery Life:
Low Battery
"Low Battery" icon displayed on LCD
Indication:
Dimensions:
3.16" x 5.10" x 1.04" (80.3 mm x 129.5
mm x 26.4 mm)
Weight:
0.4 lbs (181 grams)
CLASSIFICATIONS
Protection against
Internally powered equipment
electrical shock:
Protection against
water:
Mode of operation:
Continuous
Sterilization:
See 7.1 Cleaning, page 14.
Flammable
Not for use in presence of flammable
anaesthetic mixture:
anaesthetic mixtures
Power specification:
1.8 - 3.2 V
Technical Data
0.1 up to 199, 1
from 200 to 344
32 mW 10mA
15