Supplement for Infinity CentralStation Wide VG2.n and M300 VG1.n
New features
The next sections contain new features related to
the ICS VG2.n release.
Caliper measurement
The following enhancements now show on the
Calipers page:
A
B
D
E
A Lead button – Selects the lead to provide mea-
surements.
B Size button – Selects the size of the waveform.
C Zoom in/out
– Zooms in or out of a partic-
ular waveform to view more detail.
D Add row button – Creates a new row to contain
a new set of measurements. You can only mea-
sure one lead at a time (in one row).
E Delete row button – Deletes the current mea-
surement row.
4
C
F
Supplement Infinity CentralStation Wide VG2.n / M300 VG1.n
English
F HR (RR) column – Displays the heart rate in
bpm. The calculation is: HR (RR) averaged =
60/RR averaged if RR is in seconds, or =
(60*1000)/RR averaged if RR is in milliseconds.
System configuration
System setup - Recorders/ Reports page
The new option Preview reports allows the user to
view reports in PDF format before printing. The set-
ting made on the Recorders/ Reports page be-
comes the default value in bedview.
The new option Report hospital name allows the
user to enter a hospital name to be printed on a re-
port.
Trends/Data dialog window
Trends graph, Trends table, Full disclosure,
Events
From the Reports... button, the new selection Pre-
view reports is available. When set to On, select-
ing a report will display a PDF preview.
To toggle the Preview reports option on or off:
 Select the arrow button
To print a report from PDF preview:
 Select the
icon.
Alarms
Overview of alarms
Visual alarm message banners are intended to be
read by the user at a distance of up to 1 meter (ap-
proximately 3.3 feet) from the Infinity CentralStation
display. Flashing parameter box signals are intend-
ed to be recognized by the user at a distance of up
to 4 meters (approximately 13 feet).
Alarm functionality should be verified by generating
a high grade, medium grade and low grade alarm
condition and confirming visual and audible alarm
annunciation at the Infinity CentralStation.
.