3. Specifications
3.9 Accuracy data model O-10 (5)
Non-linearity (per IEC 61298-2)
≤ ±0.5 % of span BFSL
Measuring deviation of the zero signal
≤ ±1 % of span
Accuracy at reference conditions
≤ ±2.0 % of span
Temperature error at 0 ... 80 °C
Mean temperature coefficient of zero point:
■
Typical:
0,3 % of span/10 K
Maximum:
0,6 % of span/10 K
Mean temperature coefficient of span:
■
≤ ±0.1 % of span/10 K
Long-term drift
≤ ±0.2 % of span/year
WIKA operating instructions pressure sensor, model O-10
EN
15