Service and Maintenance

DANGER
Warning of electric shock!
Danger to life!
Switch off the machine and pull out
the mains plug before performing any
care or maintenance work.
Do not spray the machine with water
or immerse it in water.
Replacing the BEGO full demineralisation cartridge insert
30
WARNING
Unit is under pressure!
Switch steam cleaner off and allow
pressure to drop to zero!
Ensure that the pressure gauge
reads 0 bar or less before continuing.
Do not open the unit until the
pressure has dropped to zero!
1.
Turn tap (1) off.
2.
Unscrew hose (2) and allow water to drain.
3.
Unscrew casing (3) with ring spanner (see
scope of supply of BEGO full demineralisation
cartridge).
4.
Replace insert (4).
Ensure the seals are correctly positioned.
1 Insert (new) ................................. REF 12260
2 Inserts (new) .............................. REF 37602
5.
Screw casing (5) on again tightly using ring
spanner.
6.
Turn tap (6) on, allow 2 or 3 litres of water to
flow through, turn tap off.
7.
Screw hose (7) on tightly.