English ⋅ Triton SLA
en
The Triton SLA is fitted with several safety devices:
•
If the permissible operating pressure is exceed-
ed the supply of water is interrupted and the
heating switched off (switching range
3.5 ± 0.2 bar).
•
A safety valve opens at 4.4 bar.
•
If the boiler is heated without any water, a tem-
perature limiter switches the heating off.
Scope of delivery and Accessories
NOTE
The spare parts list is appended.
When ordering spare parts please quote the REF,
quantity and equipment serial no.
Scope of delivery
Triton SLA
200...240 VAC, 50/60 Hz ...................... REF 26005
100...120 VAC, 50/60 Hz ...................... REF 26006
with:
•
Key for drain cock
•
Spare fuses
•
Operating instructions
•
BEGO full demineralisation cartridge
(accessory)
Technical Data
Height..........................................................540 mm
Width ...........................................................380 mm
Depth ..........................................................280 mm
Weight .............................................................13 kg
Nominal voltage .............. 200...240 VAC, 50/60 Hz
Special voltage ................ 100...120 VAC, 50/60 Hz
26
•
A water leak detector, which is placed on the
floor under the cleaner, measures electrical
conductivity (µS) and switches the water supply
off automatically in the event of a leak
(µS indicator light comes on).
•
If the full demineralisation cartridge is being
used, the supply of water is interrupted when its
resin is spent (µS indicator light comes on).
Accessories
BEGO full demineralisation cartridge,
with 2 inserts and ring spanner ..................... 37600
Insert for full demineralisation cartridge
with seal ring:
1 piece (new)................................................. 12260
1 piece (recycled) ....................................... 12260 T
2 pieces (new) ............................................... 37602
2 pieces (recycled) ..................................... 37602 T
Durox non-reusable ion-exchanger resin, 6 l 52121
Ring spanner ..................................................11044
Calex ............................................................. 52125
Power consumption at nominal voltage........ 1.5 kW
Boiler temperature at pressure of 3 bar........ 133 °C
Steam pressure ............... 3 ± 0.2 bar (ca. 0.3 MPa)
Boiler capacity .................................................. 2.9 l
Water supply...................................... 3/4", 4...6 bar
Noise emission ...................................... < 70 dB (A)