Liquids sterilization: do not open the valve when it cools down till the pressure
gauge (1) indicates 0 bar. Then, set the valve (12) to the «Steam» position.
In both cases, please open the autoclave lid when the pressure gauge indicates 0
bar, as explained in chapter 8.2.
9 Alarm indicators
Overtemperature indicator (2). The safety thermostat has disconnected the heater:
Possible cause 1:
-
The equipment has been switched on without or with not enough water.
Possible solution:
1. Add water untill the heater element cover rack, as indicated in chapter 8.1.
4
5
2. Press the safety thermostat restart button (19) and begin the sterilization process
once again.
Possible cause 2:
-
The valve is not closed (12) and after the purge phase, the water in the sterilization
chamber has been evaporated.
Possible solution:
1. Do the same as in cause 1 and be sure, when initiating the sterilization process,
9
10
to set the valve (12) to the «Closed» position after purging. See chapter 8.3.
Open lid indicator (3)
Possible cause:
-
The indicator is on while the lid is open and must be off when closing it. If it does
not switch off or on when operating, carefully depressurize the autoclave and
contact an authorized technical service of JP Selecta, SA.
10 Spare Parts
07001
13007
15480
15595
16078
16212
16220
20061
20096
22003
22004
22063
34057
34202
37025
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Junio 2012
(Sujetas a modifi caciones sin previo aviso)
Network connection cable
Restart and heater contactors
Fuse carrier
Fuse
Pressure gauge
Pressure switch 2 bar (134ºC)
Pressure switch 1 bar (121ºC)
Temperature selector
Main switch
Amber indicator light
Red indicator light
Green indicator light
Three channels valve
Safety valve
Timer