LNC MP Series Patient Cables
Reusable
Prior to using this Cable, the user should read and understand the Operator's Manual for the Device, this Directions
for Use and the Directions for Use for the Sensor.
INDICATIONS
The LNC MP Series Patient Cables and Masimo Sensors are indicated for the continuous noninvasive monitoring of functional
oxygen saturation of arterial hemoglobin (SpO
infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
DESCRIPTION
LNC MP Series Patient Cables and Masimo Sensors have been validated with the Masimo SET and Philips FAST-SpO
listed above. The saturation accuracy of the neonatal sensors were validated on adult volunteers and 1% was added to
account for the properties of fetal hemoglobin. The LNCS YI multisite sensor was validated on adult digits.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to
use Masimo sensors.
WARNINGS
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route cable and patient cable to reduce the possibility of patient entanglement or strangulation.
• Do not attempt to reprocess, recondition or recycle Masimo patient cables as these processes may damage the
electrical components, potentially leading to patient harm.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Refer to the specific sensor's Directions for Use for detailed sensor instructions and warnings.
• Probes and cables are designed for use with specific monitors. The user and/or operator needs to verify the compatibility
of the monitor, probe, and cable before use, otherwise patient injury can result.
INSTRUCTIONS
A. Attaching LNC MP Patient Cables to the Device
Refer to Fig. 1a and 1b. Orient the connector of the cable to mate with the patient cable connector on the SpO
(Fig. 1a) or the MMS (Fig. 1b).
CAUTION:
ENSURE YOU ARE CONNECTING A LNC MP PATIENT CABLE TO EITHER A MASIMO SET OR PHILIPS
FAST SPO
AND DIFFERENT MECHANICAL KEYING. NEVER FORCE A PATIENT CABLE CONNECTOR INTO THE
MODULE OR MMS. FAILURE TO USE A LNC MP PATIENT CABLE MAY RESULT IN DAMAGE TO THE
MODULE OR MMS, INACCURATE READINGS, OR NO READINGS.
Note:
There is a difference in color shade between the LNC MP Patient Cable and the Philips MMS/SpO
receptacle. However, this is an acceptable configuration. Masimo SET performance is not available when
connected to an instrument with Philips FAST SpO
B. Attaching the LNC MP Patient Cable to a LNCS Sensor
1. Refer to Fig. 2. Insert the Sensor cable connector completely into the LNC MP Patient Cable.
2. Refer to Fig. 3. Completely close the protective cover.
C. Disconnecting the LNC MP Patient Cable from a LNCS Sensor
1. Refer to Fig. 4. Lift the protective cover to gain access to the sensor connector.
2. Refer to Fig. 5. Pull firmly on the sensor connector to remove from the patient cable.
D I R E C T I O N S F O R U S E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
) and pulse rate (measured by an SpO
2
SOCKET. THERE ARE A NUMBER OF SIMILAR CONNECTORS WITH DIFFERENT COLORS
2
Technology.
2
3
sensor) for use with adult, pediatric,
5%
2
Title: Graphics, Sensor/Cable symbols, 03/06
GR-14231
DRO-13914
4223F-eIFU-0617
en
+70 C
-40 C
+1060 hPa - +500 hPa
795 mmHg - 375 mmHg
Non-sterile
NON
STERILE
5%-95% RH
LATEX
95%
%
devices
2
Module
2
Module
2