Implementation of the provision Nitrite with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
Fill 24 mm vial with 10 ml
sample.
Zero
Press the ZERO button.
For devices that require no ZERO measurement , start here.
Add NITRITE LR tablet.
29
Nitrite T / M270
10 ml
Close vial(s).
Remove the vial from the
sample chamber.
Crush tablet(s) by rotating
slightly.
Place sample vial in the
sample chamber. • Pay at-
tention to the positioning.
24 mm
Close vial(s).
EN   Method Reference Book   1.0
Sample