WhAt doeS
eRRoR
Air Pressure Sensor: The
Control Unit has detected a
problem calibrating the air
pressure circuit on start-up.
Self-test error – Air
Pump: The Control Unit
has detected an electronic
problem in the air circuit on
start-up.
dry Pump: The Control Unit
has detected a dry pump.
In order to prevent possible
damage to the fluid pump,
the unit will stop therapy.
over Pressure: Indicates
that the Control Unit has
exceeded the target air
pressure.
t R o u B l e S h o o t i N G
it MeAN?
• Do not make sudden movements during
WhAt cAN
i do?
This is most likely to occur if you have
restarted the System with an inflated Wrap
attached.
Detach the Wrap, pressing it flat to expel the
air accumulated inside and try again.
Disconnect the Wrap from the Control Unit.
Turn the Control Unit off and on again
without a Wrap connected.
Reconnect the Wrap and resume treatment.
If the problem persists, contact Customer
Service.
Be sure you are using ATX Series Heat
Exchangers.
Note that if you are using new Wraps,
the water in the reservoir may have been
depleted and causing this error. Make sure
there is adequate water in the reservoir
based on the fill line indicator label. Refer to
instructions for priming a Wrap below.
Verify that the ice box filter is not clogged
(refer to filter maintenance instructions on
page 12).
Make sure there are no kinks in the Wrap or
Connector Hose.
Turn the Control Unit off and on again.
Disconnect and reconnect the hose from the
Control Unit and the Wrap, verifying that an
audible "click" is heard at both connection
points.
Prime the Wrap using the following steps:
- Select "No Pressure. "
- Attach the hose to the unit and the wrap.
- Lay the Wrap open and flat next to or
lower than the Control Unit (not on the
body).
- Run the System for 2 minutes.
Prime the Control Unit using the following
steps:
- Disconnect the hose from the Control Unit.
- Now, look at the Wrap connection location
on the Control Unit. On the top valve, push
the white prong in so that it is flush with
the metal connector.
- Make sure you are not fully covering the
opening in the prong.
- Push start, and water should squirt out of
the valve.
- Restart the System.
Turn the Control Unit off and back on.
Make sure the Wrap is applied securely.
treatments. Rapid shifting in position may
produce a quick change in pressure in the
Wrap and cause this error.
EN-21