Game ready GRPro 550550-03-RN Manual De Instrucciones página 47

Ocultar thumbs Ver también para GRPro 550550-03-RN:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
table 2 for transient electromagnetic immunity
The GRPro 2.1 is intended for use in the electromagnetic environment specified below. The customer
or the user of the GRPro 2.1 should assure that it is used in such an environment.
immunitY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the AC mains voltage prior to application of the test level.
T
EN-18
GuidANce ANd MANuFActuReR'S declARAtioN –
electRoMAGNetic iMMuNitY
iec 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
compliAnce level
electromAgnetic environment – guidAnce
± 6 kV contact
Floors should be wood, concrete or
± 8 kV air
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
± 2 kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
environment.
± 1 kV for input/
output lines
± 1 kV line(s) to
Mains power quality should be that
line(s)
of a typical commercial or hospital
environment.
± 2 kV line(s) to
earth
<5% U
Mains power quality should be that
T
(>95% dip in U
of a typical commercial or hospital
)
T
environment. If the user of the
for 0.5 cycle
GRPro 2.1 requires continued operation
40% U
T
during power mains interruptions, it is
(60% dip in U
)
T
recommended that the GRPro 2.1 be
for 5 cycles
powered from an uninterruptible power
70% U
T
supply or a battery.
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Tabla de contenido
loading

Este manual también es adecuado para:

Grpro 550550-53Grpro 550550-03

Tabla de contenido