The parameter configuration for the SCE-020-02 is ended as soon as
the value for Li 0 is confirmed with SET . The saving of this
configuration follows by selecting the menu End . A further confirmation
with the SET key saves the parameter configuration. The instrument
switches into the operating mode Measuring and Display .
Parameter Configuration of the SCE-020-02
1.
Set the decimal place – either decimal comma or
decimal point – according to Section 4.2.1. (watch the
input signal and measurement range)
2.
Setting the measurement ranges in 0 and in FS
Select either M3 or M4 with the + / - keys and activate
the parameter selection with SET . Then enter the
beginning and end values with the + / - keys and confirm
with SET . Pressing the SET button ends the setting.
3.
Configuration of the Baud Rate
Watch out for the compatibility between the sending and
receiving units.
4.
Scanning rate
Set the desired scan rate. Measurement readings query
is possible on demand.
Configuration of the limit values Li0 and Li1
5.
Relay contact according to operating current principle.
- Li0 < Li1
for example, limit value hysteresis = 10% ( Li1 = 110 bar,
Li0 = 100 bar)
6.
Saving the parameter configuration
The parameter configuration is ended as soon as the
value of Li1 is confirmed with SET . The saving of this
configuration is effected with End . With a further pressing
of the SET button, the parameter configuration is saved.
The instrument switches automatically into the operating
mode Measuring and Display.