Introduction

This digital multimeter complies with international safety standards
EN61010–1 electrical safety requirements for electronic measuring
instruments and handheld digital multimeters; CAT III 600V and pollution
degree of 2. Read all instructions carefully before using the meter and follow
all relevant safety standards.

Safety guidelines

Safety category Information
• Category III mesurements (CAT III) are measurements done inside the
infraestructure of a building.
Example: fixed equipment switchboards, circuit breakers, wiring, including
cables, bus bars, junction boxes, switches, sockets, output terminals on
devices for industrial use and other equipment.
• Category II measurements (CAT II) are measurements done between the
wall socket and a device.
Example: measurements on household appliances, portable tools and
similar equipment.
• Category I measurements (CAT I) are measurements done at the
component level.
Example: surface mounted devices on a circuit board.
Precautions
• Allow the meter to warm up for 30 sec. before use.
• Do not place meter in a strong magnetic field, this may cause false redings.
• Do not place the meter in any environment with high pressure, high
temperature, dust, explosive gas or vapor.
• Inspect the case before use. Check for cracks in the casing and the
insulation around the input sockets.
• Only use the test leads provided with the meter. If leads are damaged or
need to be replaced, use similar leads with matching specifications.
• Ensure the meter works properly by testing a known voltage source first. If
not working properly, the protective equipment my be damaged; have the
meter serviced before using.
• Make sure the test leads are in the correct input jacks before measurement.