Transmitter Board
4
3
2
1
Figure 5 HMD65 Transmitter Board: Service Port, DIP switches, Trimmers, and Screw Terminals
1
RS-485 termination (120 Ω resistor) ON/OFF switch.
2
RS-485 (Modbus/BACnet) screw terminals.
3
Power supply input (15 ... 35 VDC or 16 ... 24 VAC) screw terminals.
4
Service port for MI70 hand-held indicator and Insight PC software cable connection.
5
DIP switches for setting the HMD65 Modbus RTU or BACnet MS/TP MAC address.
6
DIP switches for selecting Modbus/BACnet communication bit rate and parity (Modbus
only).
7
DIP switches for humidity output parameter selection.
8
DIP switch for selecting either Modbus or BACnet mode.
9
Trimmer for humidity measurement adjustment.
10
Screw terminals for humidity measurement output.
11
Screw terminals for temperature measurement output.
12
Trimmer for temperature measurement adjustment.
13
Indicator LEDs: flash when there is RS-485 transmit (TX) or receive (RX) activity.
10
5
6
7
8
13
9
10
11
12
M212264EN-B