Time controlled
Internal start via a
timer.
▼▲
▼▲
NOTE
Selecting the operating mode
Under the menu item
the type of analysis controller. Testomat 2000
selection options: Time control, quantity control via water meter,
dynamic analysis start and external analysis start.
Shortest interval = 0 minutes between analyses. Largest interval = 99
minutes.
The analysis interval (time between two analyses) is determined by
the duration of the supplementary program AUX, the set flush times
(internal and external), the programmed interval and the duration of
the analysis. The analysis duration depends directly on the
measured value.
Selecting the time control
In
=>
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
("TIME CONTROLLED" " * " is preset)
Enter the interval pause and the flush times
In
=>
=>
Enter the "
(1 minute is preset)
Enter the "
(00 seconds (s) is preset)
Enter the "
(00 seconds (s) is preset)
Press "ENTER" to complete all the entries
Duration of the analysis interval
The analysis interval is the addition of the "AUX
before/after analysis", "Internal flush" and "External
flush" times and the measuring value dependent
analysis duration (see diagram on the left)
Password protection and basic program
, select
=>
, select
=>
in minutes (m)
in seconds (s)
in seconds (s)
it is possible to select
®
provides numerous
=>
Analysis
interval
Sequence of
times