Displayed message (flashes at
selected display)
Mf. ANALYSIS
CANCEL WITH HORN KEY
Ff. DOSING ERROR
CANCEL WITH HORN KEY
MAINTENANCE INTERVAL
EXCEEDED BY
XXX DAYS
CANCEL WITH HORN KEY
Abbreviations: Ff.: = function fault, Mf. = Measuring fault
Error
Current interface functions incorrectly
Unit is not functioning, even though it is
switched on
No display
Instrument result functions
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses or no message
- LED and output
"maintenance" on
- Continue measurements
- After programming:
Continuous alarm or message
impulses or no message
- LED and output
"maintenance" on
- Continue measurements
Further information
Possible causes
- Incorrect measured value at the output or
no power supplied
- Fuse F9, F5 or F2 (240 V: F1) defective
- Power switch defective
- Ribbon cable at display circuit board or
base circuit board is loose
- Error at display circuit board or base
circuit board
Tripping of a protective circuit
After a protective circuit (fuse) has been tripped, attempt to eliminate
the cause of malfunctioning (e.g. replace a defective valve) before
reactivating the protective circuit. Frequent tripping is usually due to
power overload which, in certain circumstances, may also damage
the instrument.
Malfunctioning/Repairing a defective instrument
The repair of a defective instrument – irrespective of the warranty
period – is only possible when the instrument is dismantled and
returned to us with a description of the error. Furthermore, please
inform us of the indicator type being used and the measured medium.
Before you return the instrument for repair work, remove the bottle
and ensure that the measuring chamber has been flushed out and is
empty.
Error messages / Troubleshooting
Possible causes
- Air inside the dosing hoses
- Incomplete mixing
- Indicator out of date or use
of a third-party indicator
- Dosing accuracy of the
dosing pump
- Programmed maintenance
date reached or exceeded
Remedies
Replace fuse F7
Replace the interface circuit board
Replace fuses
Replace power switch
Reconnect ribbon cable
Replace display or base circuit board
Remedies
Retighten dosing pump
connections
Replace intake insert in
the bottle
Replace stirring bar
Replace indicator, only
use HEYL Testomat
®
2000
indicator
Replacing dosing pump or
return it for calibrating
Carry out maintenance
and subsequently cancel
or confirm
49