Switch functions of the limit value outputs LV1 and LV2

Switch functions of the limit value
outputs LV1 and LV2

Switch function 0, duration

If the limit value LV1 or LV2 has been exceeded, the output relay LV1
or LV2 reacts. If the measured value falls below the limit value LV1
or LV2 without locking, the relevant relay drops out again.

Switch function 1, impulse

If the measured value exceeds the limit value LV1 or LV2, the
relevant output switches for a settable time (t).
The respective output always remains switched for the set time,
irrespective of how long the limit value has been exceeded. A new
impulse is only possible once the value has fallen below the limit
value!

Switch function 2, interval

If the limit value has been exceeded, the respective output switches
at intervals with the settable time (t) = impulse or interval as long as
the limit value is exceeded. The switching on time and the interval are
the same.

Switch function 3, two-point

If the upper limit value LV1 has been exceeded, the output relay LV1
switches. If the value falls below the lower limit value LV2, the output
relay LV1 drops out again. The output relay LV2 switches according
to the programmed switch function.
In
=>
Select the duration, impulse, interval or two-point (only for LV1)
Enter the time (only for switch function 1 and 2)
Press "ENTER" to confirm the entry
The function 3 is only possible if different values are used for the limit
values LV1 and LV2 with just one measuring point. For example, for
LV1 = 0.2 °dH and for LV2 = 0.1 °dH.
Switch functions and locking
Switch functions 0 and 2: If locking has been programmed, the
output relay LV1 switches as programmed until manually released
(press the "STANDBY" key).
If locking has been programmed, switch function 3 cannot be
selected!
, select
=>