Service II

▼▲
The service II menu contains various functions for monitoring the
operation of the instrument:
Programming of the maintenance interval, operation (reset) internal
data/setting, e.g. water quantity and plant monitoring.
Use of the Service II menu
NOTE
The functions in the Service II menu directly influence the operation
and the monitoring functions of the instrument and can result in plant
malfunctioning!

Reset operating time

After replacing a dosing pump or the measuring chamber holder, it is
possible to reset the current operating time to 0 hours.
▼▲
The operating time "000000h" appears on the display

Maintenance interval

Observance of the maintenance intervals is monitored and displayed
by Testomat 2000
days here. (0 days equals no maintenance interval)
▼▲
▼▲

Water quantity RESET

Causes a reset of the currently stored water quantity. This has a
direct effect on the plant monitoring and the dynamic analysis control.

Plant control RESET

Plant monitoring is only re-activated when the limit value is exceeded
again.
Switch functions of the limit value outputs LV1 and LV2
These tasks should only be carried out by trained and qualified
staff.
In
, select
=>
=>
Select "Reset" to reset the operating time
Press "ENTER" to confirm the selection
®
. Program the desired maintenance interval in
In
, select
=>
=>
Enter the maintenance interval in days (d)
Press "ENTER" to confirm the entry
In
, select
=>
=>
Press "ENTER" to confirm the entry