Comparing the Control Solution Test Results

The test result of each control solution should be within the range
printed on the label of the test strip box. Repeat the control
solution test if the test result falls outside of this range. Out of
range results may occur in the following situations:
When the control solution bottle was
not shaken well,
When the meter, test strip, or the
control solution were exposed to high
or low temperatures,
When the first drop of the control
solution was not discarded or the tip
of the bottle was not wiped clean,
When the meter is not functioning
properly.
When the control solution is past the
expiration date printed on the bottle,
When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
When the control solution is
contaminated.
If results continue to fall outside the range printed on the test
strip box, the KetoSens Test Strip and Meter may not be working
properly. Do not use your system and contact Customer Service:
1-800-429-5001 (9 am–9 pm EST, Monday–Saturday). At all other
times, contact your healthcare professional.
Situations
Do This
Repeat the control
solution test by
referring to the Note
on page 23.
Discard the used
control solution
and repeat the test
using a new bottle
of control solution.
www.i-sens.com 27