Comparing the Control Solution Test Results
The test result of each control solution should be within the range
printed on the label of the test strip box. Repeat the control
solution test if the test result falls outside of this range. Out of
range results may occur in the following situations:
When the control solution bottle was
•
not shaken well,
When the meter, test strip, or the
•
control solution were exposed to high
or low temperatures,
When the first drop of the control
•
solution was not discarded or the tip
of the bottle was not wiped clean,
When the meter is not functioning
•
properly.
When the control solution is past the
•
expiration date printed on the bottle,
When the control solution is past its
•
discard date (the date the bottle was
opened plus three (3) months),
When the control solution is
•
contaminated.
If results continue to fall outside the range printed on the test
strip box, the KetoSens Test Strip and Meter may not be working
properly. Do not use your system and contact Customer Service:
1-800-429-5001 (9 am–9 pm EST, Monday–Saturday). At all other
times, contact your healthcare professional.
Situations
Do This
Repeat the control
solution test by
referring to the Note
on page 23.
Discard the used
control solution
and repeat the test
using a new bottle
of control solution.
www.i-sens.com 27