6.
Using the foot switch of the HF device, activate the HF
current for coagulation:
A continuous acoustic signal sounds during
the entire sealing process to indicate that
power is being supplied.
An alternating acoustic signal indicates the
end of the sealing process.
7.
Release the foot switch to stop the supply of power.
8.
Using the ratchet lever C1, open the jaws A1.
The tissue is coagulated.
Cutting tissue (only with NightKNIFE)
WARNING
Strong bleeding may occur if the grasped tissue is cut
before coagulation or ligation!
With blood vessels, place at least two seals to the left
and right of the cutting site.
Before cutting, ensure that the tissue has been reliably
sealed.
Cut only in the sealed area.
The tissue is grasped by the jaws and is coagulated.
1.
To cut, pull the trigger B3 and then release it.
The tissue is cut.
2.
Using the ratchet lever C1, open the jaws A1.
NightKNIFE / Ligator
5 Operation
25