6556042.book Seite 32 Freitag, 15. April 2016 3:13 15
7 Conditioning
32
7.2
Automated cleaning and disinfecting
BA 755 / BA 758 can also be cleaned and disinfected in suitable
Product name
cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ BA 755 / BA 758 is conditioned with a cleaning and disinfection
Check _ product name
device.
1. Check whether BA 755 / BA 758 is clean after processing.
2. If it is still dirty, repeat the process.
For further processing, BA 755 / BA 758 should be dry and free
of residue.
Operating Instructions BA ULTIMATE POWER+
65 56 042 D3560.201.08.02.02