Feed & Flush Pump Sets; Re-Priming After Bag Empty; Selecting Feed Mode; Ez Pump Mode Option - Covidien Kangaroo Joey Manual De Instrucciones

Bomba para irrigación y alimentación enteral joey con abrazadera para palo de gotero, programable
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 192
Feed & Flush Pump Sets
For feed & flush pump sets, the display will appear as in Figure 8B. First press and hold
flushing line, and then press and hold
end of the pump set.
If a feed & flush pump set is used, remember to first press
and then press
HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the stepped end connector.
If the feeding line is primed first, subsequent priming of the flushing line will force the air between the flush solution bag and valve
into the main line, which already has been primed with enteral formula.

Re-priming after Bag Empty

A pump set bag that has been emptied will trigger the FEED ERROR screen. In this condition the pump set bag can be refilled to
continue the feeding, but only after the pump set has been re-primed.
To re-prime the pump set, do the following:
• Disconnect the feeding line from the patient
• Refill the bag
• Press
CONTINUE to begin the pump RUNNING
HOLD
• Press
• Press
ADJUST SETTINGS
PRIME PUMP
• Press
• Press
HOLD TO PRIME to prime the line(s)
Pump sets should not be reused after 24 hours of initial usage. The feeding set should also be replaced after 24 hours from initiation
of feeding. This ensures that the system is operating within specified parameters and prevents bacterial growth that could be a
hazard to the patient.

Selecting Feed Mode

EZ Pump Mode Option

The EZ PUMP MODE option is a limited, CONTINUOUS MODE only setting of Kangaroo™ Joey enteral feed and flush pump with
pole clamp. It feeds non-stop (no VTBD option) at a rate 0-400mL/hr as defined by the user until manually stopped. It has no
BOLUS, FLUSHING, RUN MODE SCREEN LOCK, or AUTORESUME capability and displays no HISTORY information.
Turning on EZ PUMP MODE will automatically force the unit out of the INTERMITTENT (BOLUS FEED) MODE if set. Although the
Continuous Mode feed VTBD will be set to 0 while EZ PUMP MODE is enabled, the previous value of the FEED VTBD will be stored
in the pump memory and recalled when "EZ PUMP MODE" is deselected. The user can change the Feed Rate by normal means. All
other user settings, including those not compatible with the EZ PUMP MODE (BOLUS VTBD, NUMBER OF BOLUSES, SUPER
BOLUS, etc), that were set before the pump was put into the EZ PUMP MODE will be ignored and invisible while in EZ PUMP
MODE but remain stored in the pump memory for when EZ PUMP MODE is deselected. EZ PUMP MODE essentially has no effect
on CLEAR SETTINGS; pressing CLEAR SETTINGS while EZ PUMP MODE is enabled will clear all feed and flush settings and
totalizers as normal, not just those that apply to EZ PUMP MODE. Selecting the CLEAR SETTINGS option on power up will not
clear Biotech Options such as EZ PUMP MODE or LOCK SETTINGS.
If the user selects the ADJUST FEED option from either the LOAD A SET or SET LOADED screens, control will pass directly to and
from the Enter Feed Rate screen and bypass the ADJUST FEED screen since without the ability to adjust VTBD in EZ PUMP MODE,
the Adjust Feed screen is redundant.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
HOLD TO PRIME FEED to prime the feed line down to the stepped end connector at the
HOLD TO PRIME FLUSH until the flushing fluid has reached the valve,
12
HOLD TO PRIME FLUSH to prime the