Guidance and manufacturer's declaration – electromagnetic immunity for not life
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Immunity
IEC 60601
tests
Test level
Radiated RF
3 V/m
IEC
80 MHz to
61000-4-3
2.5 GHz
162/172
support equipment
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to
any part of the Product, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 0.35 √P
d = 0.7 √P
Where P is the maximum output power
rating in the transmitter in watts (W)
3 V/m
according to the transmitter
80 MHz to
manufacturer and d is the recommended
2.5 GHz
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
less than the compliance level in each
frequency range
Interference may occur in the vicinity of
equipment marked with the following
symbol: