Step3 is the process of warming the chamber
o
temperature to 45
C.
Note: By pressing "Abort", decontamination will be
stopped and the procedure jumps to Step8 (end of
decontamination).
Step4 is the process of decontaminating the
chamber by generating H
generator.
Note: By pressing "Abort", decontamination will be
stopped and the procedure jumps to Step7
(resolution of H
O
vapour).
2
2
Step7 is the process of resolving H
using the UV lamp.
Note: Step5 and Step6 do not exist.
15. When H
O
decontamination is completed, the
2
2
H
O
Decontamination Step8 screen is displayed.
2
2
Open the outer and inner doors. Disconnect the
connector, cable, and H
O
2
the H
O
generator and cable from the chamber.
2
2
Note: Put on protective glasses and rubber gloves
to do this.
O
vapour from the H
O
2
2
2
2
O
vapour
2
2
generator, and remove
2