CORRECT OPERATION
Use the following procedure to start trial operation or actual operation of the incubator.
1. Install the incubator correctly, referring to pages 17–20.
2. Remove the packing materials from the chamber and inner attachments. Clean and disinfect the
chamber and all the inner attachments, referring to page 21.
3. Add approximately 1.5 L of sterile distilled water to the humidifying pan. (Refer to page 25.)
4. Connect the removable power supply cord to the port on the lower right side.
Note: Two removable power supply cords are provided.
For UK
For EU countries other than UK
5. Connect the removable power supply cord to the outlet.
6. Turn ON the power switch on the lower right side of the incubator.
7. (For model MCO-230AICUV or when an optional UV system set MCO-170UVS is installed) Set the
frequency of power supply on the LCD touch panel (Refer to page 48).
The removable power supply cord provided is only for this product. Never use it for any other products.
28