Cleaning, Disinfection and Sterilization
Operation
9
Sterilization
10
Storage
30/222
Operating mode
•
Steam sterilization at:
134 °C / 273°F,
3 min.
•
Keep devices in
sterilization
packaging in a dry
and clean
environment.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Warning
•
The accessories (lip clip,
hook and fork) must be
sterilized according to the
packaging labelling.
•
Use only autoclaves that
match the requirements of
EN 13060, EN 285.
•
Respect the maintenance
procedure of the autoclave
device given by the
manufacturer.
•
Use only this recommended
sterilization procedure.
•
Control the efficiency
(packaging integrity, no
humidity, colour change of
sterilization indicators,
physico-chemical
integrators, digital records of
cycles parameters).
•
Sterility cannot be
guaranteed if packaging is
open, damaged or wet.
•
Check the packaging and the
medical devices before using
them (packaging integrity, no
humidity and validity period).