Usb Data Recorder - Selecta Autester ST Dry PV III Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CÓDIGO 80311 REV E
************************
14/05/2015
09:52
========================
PROG: 0
MODE: SOLID2
TEMP: 134°C
TIME: 04'
DRY : 10'
========================
LOT: 000010
************************
TIME TEMP. PRESS. STATUS
023°C +0.0bar VACUUM
021°C -0.2bar VACUUM
021°C -0.4bar VACUUM
021°C -0.5bar VACUUM
022°C -0.5bar
HEAT
023°C -0.5bar
HEAT
023°C -0.5bar
HEAT
025°C -0.5bar
HEAT
060°C -0.3bar
HEAT
071°C -0.1bar
HEAT
080°C +0.0bar
HEAT
089°C +0.2bar
HEAT
095°C +0.3bar
HEAT
100°C +0.5bar
HEAT
095°C +0.0bar VACUUM
090°C -0.1bar VACUUM
086°C -0.3bar VACUUM
083°C -0.4bar VACUUM
081°C -0.4bar VACUUM
078°C -0.5bar VACUUM
076°C -0.5bar VACUUM
079°C -0.4bar
HEAT
088°C -0.2bar
HEAT
095°C +0.0bar
HEAT
101°C +0.1bar
HEAT
106°C +0.3bar
HEAT
110°C +0.5bar
HEAT
114°C +0.7bar
HEAT
119°C +0.9bar
HEAT
122°C +1.2bar
HEAT
126°C +1.4bar
HEAT
129°C +1.7bar
HEAT
133°C +2.0bar
HEAT
134°C +2.0bar
HEAT
03' 134°C +2.0bar STERIL
02' 134°C +2.1bar STERIL
01' 134°C +2.1bar STERIL
00' 134°C +2.1bar STERIL
132°C +1.9bar
EMPTY
124°C +1.1bar
EMPTY
112°C +0.4bar
EMPTY
104°C +0.1bar
EMPTY
00' 099°C +0.0bar
01' 093°C -0.2bar
02' 089°C -0.3bar
03' 085°C -0.4bar
04' 082°C -0.4bar
05' 079°C -0.5bar
06' 076°C -0.5bar
07' 073°C -0.6bar
08' 071°C -0.6bar
09' 069°C -0.6bar
068°C +0.0bar UNPRES
RESULT: PASS
Example of registration obtained by
the printer
J
.P. SELECTA s.a.u.

9 USB data recorder

The equipment has an internal memory for data acquisition (Datalogger). Once
the cycle is finished, it is possible to download this information to a USB memory
through a USB connector.
The device (USB pen drive) must be formatted to FAT 32.
Operation
• By default, the logger is active with a 1 minute logging interval. Follow the
instructions to modify the recording interval (see 8.2.7).
With the cycle completed:
• Connect the «USB pen drive» to connector.
• When you connect a "USB pen drive", the USB symbol will appear in the lower
bar of the screen.
• From the touch screen select:
• The screen will indicate when the recording of the logs has finished. Disconnect
the «USB pen drive» from the device.
• Connect the «USB pen drive» to a computer to download the data file.
Important Note: The «USB pen drive» must have a device name. Can not be
blank.
The created file is of text type, that can be opened with WordPad, WORD or EXCEL.
Note: If the data is not downloaded, when the device memory is at 80% capacity, a
warning message will appear indicating that the memory needs to be downloaded.
If the message is ignored and the memory is completely filled, the oldest cycles
will be erased.
Approximately the memory has capacity to store about 150 cycles (it will depend
on the total time that each cycle has).
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
Ene 2020
(Sujetas a modificaciones sin previo aviso)
SETTINGS> DOWNLOAD LOGS
A download button will appear and by pressing it all logs will be down
loaded and the internal memory emptied.
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pag.: 39
Tabla de contenido
loading

Este manual también es adecuado para: