6.3 Cycle types and temperature range
The autoclave allows you to configure the sterilization cycle according to the charac-
teristics of the load to be sterilized:
Purge type
With internal
Atmospheric
105°C-134°C
Vacuum 2 fractions
115°C-134°C
Vacuum 3 fractions
115°C-134°C
Atmospheric
105°C-134°C
Sterilizing time range:
Drying time range:
6.4 Pre-set cycles
Autoclave incorporates 9 pre-set working cycles which can be modified according to
the user needs.
Program
No.
0
Not wrapped load
1
Wrapped load
2
Prions
3
Delicate not wrapped
4
Delicate wrapped
5
Instrumental
6
Disinfection
7
Liquids
8
Solids
9
Cleaning program*
* Fixed program, it can not be modified.
6.5 Test cycles
Also, the autoclave Autester ST DRY PV III incorporates two test programs.
Bowie Dick test
Vacuum Test
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
(It can be modified without notice)
With external
probe
probe
105°C-130°C
105°C-130°C
From 3 to 60 minutes
From 0 to 60 minutes
Temperature
ºC
(minutes)
134
134
134
121
121
126
105
121
121
134
134ºC
0,7 bar
Page: 30
Material type
Unwrapped instruments
Bagged instruments
Double bagging, porous
Liquids, culture media
Time
Dry
Purge
4
10 2x vacuum cycles
12
30 3x vacuum cycles
20
20 3x vacuum cycles
15
15 2x vacuum cycles
30
30 3x vacuum cycles
12
25 3x vacuum cycles
25
15
Atmospheric
30
-
Atmospheric
15
30
Atmospheric
4
30
Atmospheric
3' 30''
5+ 10'