Automated Cleaning
Clean before and after each use:
1. Place the protective cap securely onto the Remote Hand Control cable connector.
2. Carry out a first soaking bath in a pH-neutral detergent such as CIDEXPLUS
of 15 minutes (or an equivalent solution) and refer to cleaning agent solution manufacturer's
instructions for immersion time and bath temperature.
3. Carry out five (5) to-and-fro motions with a swab-type brush on all surfaces, paying particular
attention to the buttons, the screws on the back side and the seam around the sides.
4. Check for remaining debris and remove as needed.
5. Carry out a second soaking bath in a pH-neutral detergent such as CIDEXPLUS for a minimum of
15 minutes while using an ultrasound bath frequency of 50–60Hz (or an equivalent decontaminant
solution) and refer to cleaning agent solution manufacturer's instructions for immersion time and bath
temperature.
6. Rinse with deionized water for 30 seconds.
7. Confirm that the protective cap is securely on the Remote Hand Control cable connector.
8. Clean in a validated washer disinfector using the "INSTRUMENTS" cycle and a pH neutral cleaning
agent intended for use in automated cleaning. Effective cleaning can be achieved using the following
parameters:
• Pre-wash at ambient temperature for 4 min.
• Enzyme washes with enzymatic cleaning agent for 4 min.
• Wash at 60°C with pH neutral cleaning agent for 3 min.
• Rinse with hot water for 20 seconds.
• Dry at 95°C for 5 min.
9. After automated cleaning, proceed to inspection instructions below.