INSTRUCTION MANUAL CODE 80315 REV B
PRG 00
Ramp UP
00:00
00:00
00:00
00:00
00:00
00:00
RPM
2000
1000
10'
25'
15'
PRG
RAMP
01
00:00
RCF
J
.P. SELECTA s.a.u.
rpm
Time
0
00:00
0
00:00
0
00:00
0
00:00
0
00:00
1h15'
35'
SET
SET
0
00:00
OPEN
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
October 2019
(It can be modified without notice)
7.4
Profile programming
To program a profile, press:
A speed profile is formed by two segments.
One segment establishes:
RAMP time to reach speed.
1.
Centrifugation speed (rpm).
2.
3.
Centrifugation time (hh:mm).
If RAMP time is zero or very small, the centrifuge will be conducted in the
quickest way.
A segment with time 0 is not executed.
The last ramp time is braking to stop the rotor.
Centrifugation profile example
PRG 01
Ramp UP
00:10
00:15
00:35
The process monitoring screen indicates at any time the ramp time, speed
(rpm) and centrifugation time.
rpm
Time
1000
00:25
2000
01:15
Tel 34 937 700 877 Fax 34 937 702 362
Page: 16