Operation
4.4 Sensor diagnostics menu
Select sensor status>Select sensor(if more than one sensor is attached)
Error list
Warning list
Note: For more information about error messages and warnings refer to
4.5 Sensor setup menu
Select sensor (if more than one sensor is attached)
Calibration
Correction factor
Offset
Offset adjustment
Standard calibration
Calibration
configuration
Reset calibration
Settings
Name
Parameter
Unit
Interval
Response time
Cleaning...
Wiper mode
6
Displays all actual error messages:
Moisture, R < M, dExt < 0.0, Wiper position is unknown, Wiper is blocked, Flash failed!, R is too
high, Shaft seal replacement, Sensor is missing.
Displays all actual error messages:
EM is too high, Concentration is too high, Calibration test., Replace the wiper., Service is due,
Seal replacement is due, Shaft seal replacement
Correction factor for the measured value.
Possible settings: 0.80–1.20
Default: Factor = 1
Adjustable from –250 to +250 mE for zero point correction
Default: Offset = 0
Perform a zero point calibration
Perform a single point calibration
Select Output mode or Calibration interval
Output mode: Select the behavior of the outputs during calibration for zero point setting (Hold,
Active, Transfer, Choice). Hold maintains the last reading prior to going into the menu. Active
transmits the current level readings, corrected with previous calibration data until new data is
entered. Set Transfer transmits the value designated during the system setup
Calibration interval: Enter number of days
The instrument resets the settings to the default configuration.
Can be edited as required (up to 10 characters)
NOx-N or NO3 (eco only NOx-N)
Unit for the measured result.
Possible settings: mg/l, ppm
eco/clear: 5, 6, 10, 12, 15, 20, 30 min
plus: 15, 20, 30 sec; 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 min
Note: Intervall of the data log is the same like the measuring intervall.
Indication of the actual response time in Counts (count x measuring interval = response time)
eco: 3–6 x interval
clear: 1–6 x interval
plus: 1–12 x interval
Note: Gliding average over 2-12 measurements.
eco, clear: 1/Measurement
plus: 1/Measurement; 1,2,3,5,6,10,12,15,20,30 min; 1,2,3,4,6,12 h, 10:00 h
Wiping interval.
Select Single or Double A-B-A or Double B-A-B
Single: Default setting (Default: eco)
Double A-B-A: Double wiping frequency
Double B-A-B: Double wiping frequency (Default: plus, clear)
Section 6 on page
17.
English