Method
Take one HVE Mirror Tip (Figure 6) and insert directly into the HVE Hose
Adapter (Figure 7).
Ensure the HVE Mirror Tip is secure in the HVE Hose Adapter.
*Before patient use, HVE Mirror Tips must be cleaned and sterilized.
Go to "Infection Control Procedures" for reprocessing instructions.
Ensure the operatory suction system is on. Refer to the manufacturer's
instructions for use.
On the chair's HVE valve, switch the lever from off to on.
Once on, suction should be flowing through the HVE Mirror Tip and ready
for use.
9
Figure
Figure 6
Figure 7

INFECTION CONTROL PROCEDURES

INSTRUCTIONS FOR REPROCESSING THE HVE MIRROR TIP
Reprocessing Step
Method
Point of Use
• Turn suction off by switching the lever from on to off on existing
HVE valve.
• Remove the HVE Mirror Tip from the HVE Hose Adapter.
• It is recommended that instruments are reprocessed as soon as is
reasonably practical following use.
Cleaning
• Remove gross soil by rinsing under water of at least drinking quality.
• Prepare enzymatic instrument cleaning solution per the
manufacturer's recommendation in a soaking container or ultrasonic
bath.
• Place the HVE Mirror Tip in the solution for the time specified by the
solution manufacturer for cleaning.
• Rinse the HVE Mirror Tip under running water of at least drinking
quality for a minimum of 30 seconds to remove detergent residue.
• Allow the HVE Mirror Tip to dry.
• Visually inspect the HVE Mirror Tip for visible soil. If visible soil is
present, repeat the cleaning procedure.
Warning
• Do not attempt to remove
the HVE Mirror Tip from the
HVE Hose Adapter while the
HVE system is operating.
• Do not use hard wire brushes
to clean the HVE Mirror Tip
as scratching can occur.
Dentsply Sirona recommends
the use of Resurge
Instrument
®
Cleaning Solution.