7 - Installation of the protective cover for the cosmetic
Pull protective cover (g) on tube (h) and let it slide upwards until stud (c1) of
mobile support (c) comes into hole (g1) of protective cover (g).
8– Product life
This component has been tested in accordance with standard ISO10328, load
level P5 (100 kg, 220 lbs).
The cyclical test of 3 million cycles corresponds to a use of 4 or 5 years
depending on the patient activity.
9– Maintenance
A yearly control of the knee is recommended to check proper functioning and to
clean it (dust). If necessary, proceed to a new adjustment of the knee.
If needed, extension assist 1M11510 can be ordered separately.
10 – Advice for use, maintenance and safety (inform the patient)
In order not to damage the knee, do not use talcum powder, grease or a
silicon spray to eliminate possible rubbing noise. Any use of these
products may damage the mechanical parts, which may lead to function
defects and risk of falling for the patient.
The use of talcum powder, grease or silicon spray will render all claims
against PROTEOR null and void.
It is strictly forbidden to tighten or loose any screw of this knee except for the
4 adjustment screws R, F and S (Fig.3, 4 and 5) that can be used by the
prosthetist only.
The warranty is not applicable to damages due to improper use, inadequate
alignment, use in a very dusty environment and without adapted protection,
or any inappropriate use.
There is a risk of having a finger caught or clothing pinched in the knee. To
avoid any risk of injury caused by the moving joint, take care that no one puts
his finger near or inside the mechanism.
It should be avoided to expose the knee in environments that can cause
corrosion of metal parts (fresh water, sea water, chlorinated water, acids,
etc.)
In case of fall, shock or abnormal behaviour of the prosthesis, the knee
should be checked by the prosthetist.
The temperature range for knee use is from -10°C to +40°C (from 14°F to
104°F).
Please give a copy of these instructions for use to the patient.
11 – Recycling—End of life
This product includes items of various materials : elastomer, titanium, steel and
brass.
They must be recycled in compliance with the laws in force.