OHAUS Defender 6000 Manual De Instrucciones página 58

Ocultar thumbs Ver también para Defender 6000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

EN-56
Defender 6000 Indicators
Figure 5-3. i-DT61PW Wire Sealing
Figure 5-4. i-DT61XWE Paper Sealing
When the i-DT61PW or i-DT61XWE indicator is connected to an OHAUS Defender 6000 series base boasting
EasyConnect
TM
function
.
It has a memory module on the load cell cable, and the connection cable is sealed by
software pairing, so no hardware sealing of the cable is needed. Once the indicator or the base is replaced by a
Error 8.9
new one, an error message (
) will pop up in the indicator's displayed window.
When the i-DT61PW or i-DT61XWE indicator is connected to an outside base that does not include a memory
module, the connection between the indicator and load receptor shall be sealed by using a connector cover
(P/N: 30538022), sealing sticker or wire seal.
Figure 5-5.
Connecting cable with connector and memory module
Figure 5-6.
Connecting cable with connector and without memory module
Tabla de contenido

Solución de problemas

loading

Tabla de contenido