Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze out any
excess liquid in order to prevent damage to the meter.
2. Wipe all meter s exterior surface display and buttons. Hold the meter
with the test strip slot pointing down and wipe the area around the test
slot but be careful not to allow excess liquid to get inside. Keep meter wet
with disinfection solution for a minimum of 2 minutes for Micro-Kill+™
wipes. Follow the instructions on the package label of disinfecting wipe.
Do not use the wipe to clean or disinfect the pressure cuff.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them.
Wash your hands thoroughly with soap and water after handling the
meter, lancing device and test strips to avoid contamination. Improper
system cleaning and disinfection may result in meter malfunction. If you
have a question, please contact local customer service at 1-888-307-
8188 for assistance.
This device has been validated to withstand up to 6,000 cleaning
and disinfection cycles using the recommended disinfecting wipe/
towelette. The tested number of cycles is estimated by 5 cleaning and
disinfection cycles per day over 3 years, the expected life of the device.
The meter should be replaced after the validated number of cleaning and
disinfection cycles or the warranty period, which ever comes first.