General technical specifications
Power supply
90W power supply unit
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
30W power supply unit
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 20W (40VA)
Maximum power consumption: 36W (75VA)
90W DC/DC converter
Nominal inputs: 12V, 24V
Typical power consumption: 70W
Maximum power consumption: 110W
Operating temperature: 41°F to 95°F (+5°C to +35°C)
Environmental
Operating humidity: 10 to 95% non-condensing
conditions
Operating altitude: Sea level to 8,500' (2,591m)
Storage and transport temperature: -4°F to 140°F (-20°C to +60°C)
Storage and transport humidity: 10 to 95% non-condensing
ResMed confirms that the device/s meets the Federal Aviation Administration
Aircraft use
(FAA) requirements (RTCA/DO-160G, section 21, category M) for all phases of
air travel.
Product complies with all applicable electromagnetic compatibility requirements
Electromagnetic
compatibility
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 1 m
away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on www.resmed.com, on the Products page
under Service and Support. Click on the PDF file for your language.
IEC 60601-1
Class II (double insulation), Type BF
classification
7
English