Problem/possible cause
Device does not stop when you remove your mask
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available
if Leak Alert is enabled.
SmartStart/Stop is enabled but the device does not stop automatically when you remove your
mask
Incompatible mask system being
used.
The patient is using a nasal pillows
mask with a set pressure less than
7 cm H
O.
2
Pressure rises inappropriately
Talking, coughing or breathing in an
unusual manner.
Mask cushion is buzzing against the
skin.
Cushion seated incorrectly causing
excessive leak.
Displays error message:
Device has been left in a hot
environment.
There is a fault in your ClimateLine or
MAX
ClimateLine
heated air tubing.
Displays error message:
Device has been left in a hot
environment.
4
Solution
Enable SmartStart/Stop.
Only use equipment recommended by ResMed.
Disable SmartStart/Stop
Avoid talking with a nasal mask on, and breathe as normally
as possible.
Adjust the headgear.
Adjust headgear or re-fit cushion.
Heated tube fault, replace tube
Allow to cool before re-use. Disconnect the power cord and
then reconnect it to restart the device.
Discontinue using your ClimateLine or ClimateLine
heated air tubing and contact your clinician/service provider.
Use SlimLine, Standard or 3 m air tubing in the interim.
Humidifier fault, replace humidifier
Allow to cool before re-use. Disconnect the power cord and
then reconnect it to restart the device.
MAX