Operation; Shut Downs; Use Of Dust Filter - Dräger PEX 3000 Instrucciones Para El Servicio

Ocultar thumbs Ver también para PEX 3000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Operation

Operation
— Corresponding to the gas concentration the transmitter produces is a current bet-
ween 4 and 20 mA, especially
Output current
Meaning
0 mA
Cable is broken or power supply has failed
1 mA
Failure:
3.4 mA
Calibration signal
3.8 mA to 20.5 mA Output signal normal operation
4 mA
Output signal zero
20 mA
Output signal full scale value
20.5 mA
Overrange by more than 3 % of full scale value

Shut downs

In the event of shut-downs, e.g. when performing maintenance and inspection work,
the transmitter is ready for use again after approx. 10 minutes (sensor warm-up time)
following a renewed switch-on of the system.
Calibrate the transmitter again if necessary.
Use of dust filter
Type XTR 0000:
A dust filter can be used with the Ex-sensor PR M (see order list, part-number
68 10 537).
This is simply pressed into the opening of the sensor in front of the sinter disk and is
self-clamping. This dust filter will increase the sensor's response time only slightly.
There is nearly no influence on the sensor's sensitivity by the dust filter.
If a dust filter is used it has also to be used during the calibration procedure. It is
recommended to use a freshly installed dust filter before starting the calibration
procedure.
Page 20 of 42
Zero underrange by more than 5 % of full scale value
Failure of electronics
Broken cable or short-circuit in the sensor circuit
9023810 - 2nd edition - September 2005
Tabla de contenido
loading

Tabla de contenido