Setting Up The Transmitter - Medtronic MiniLink REAL-Time Manual Del Usaurio

Ocultar thumbs Ver también para MiniLink REAL-Time:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MP6025647-2AF1 / A
3
Within 10 seconds after the transmitter is connected, a green
light on the charger will flash for 1-2 seconds as the charger powers
on. For the rest of the charging time, the charger's green light will
continue to flash in a pattern of four flashes with a pause between
the four flashes.
4
When charging is complete, the green light on the charger
will stay on, without flashing, for 15–20 seconds and then
turn off.
5
After the green charger light turns off, disconnect the
transmitter from the charger. The green light on the
transmitter will flash for about 5 seconds and then turn off.

Setting up the transmitter

For instructions on setting up your transmitter, see the user guide for your pump.
Connecting the transmitter to the sensor
Before connecting the transmitter and sensor, make sure of the following:
•  The transmitter is fully charged.
•  Consult the user guide for your pump for details on setting up and starting the
sensor.
•  Insert the sensor and wait five minutes.
•  The sensor insertion site is not bleeding.
To connect the transmitter to the sensor:
1
After the sensor is inserted, apply the Enlite overtape to the sensor. Consult the
user guide for your Enlite serter for details on applying Enlite overtape.
2
Touch the rounded end of the inserted sensor to prevent it from moving during
connection.
3
Hold the transmitter as shown. Line up the two notches on
the transmitter with the side arms of the sensor. The flat
side of the transmitter should face the skin.
4
Slide the transmitter onto the sensor until the sensor's
flexible arms snap into the notches on the transmitter. If the
transmitter is properly connected, and if the sensor has had
enough time to become hydrated, the green light on the
transmitter will flash within 10 seconds.
-5-