Charging The Transmitter - Medtronic Guardian 2 Link Manual De Instrucciones

Tabla de contenido
MP6025871-12C1 / A
3
Slide the cover back on the charger until it clicks into place.

Charging the transmitter

Caution:
To charge the transmitter:
1
Connect the transmitter to the charger by lining it up, flat
side down, with the charger. Push the two components
together fully.
2
Within 10 seconds after the transmitter is connected, a
green light on the charger will flash for one to two seconds
as the charger powers on. For the rest of the charging time,
the charger's green light will continue to flash in a pattern of
four flashes with a pause between the four flashes.
3
When charging is complete, the green light on the charger
will stay on, without flashing, for 15 to 20 seconds and then
turn off.
4
After the green charger light turns off, disconnect the
transmitter from the charger. The green light on the
transmitter will flash for about five seconds and then turn
off.
Inserting the sensor
Always refer to the serter user guide for instructions on how to insert the sensor.
Charge the transmitter after each sensor use. A fully charged
transmitter works at least six days without recharging. A depleted
transmitter can take up to one hour to recharge.