Cavitron ® 30K 30K Ultrasonic Inserts Tm; Foot Control Information & Operation - DENTSPLY Cavitron Instrucciones Para El Uso

Ocultar thumbs Ver también para Cavitron:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Section 7: Cavitron
Cavitron Select™ SPS™ Ultrasonic Scaler Description
Cavitron
®
®
®
®
7.4 Cavitron
30K
30K
30K Ultrasonic Inserts
®
TM
TM
The many styles of Cavitron
booklet for specific information.
Insert Tip
Shape and size of tip
determines access and
adaptation. Lavage
directed to tip.
Hold the handpiece in an upright position. Activate the Foot Control to bleed any air bubbles that might be trapped inside the handpiece.
Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Fully seat insert with a gentle push-twist
motion. DO NOT FORCE.
7.5 Foot Control Information & Operation
The Foot Control is a two position momentary switch, which activates both ultrasonic energy and Lavage Flow at the insert tip.
Foot Control
released...
Both ultrasonic activation and
irrigating flow stop.
ENGLISH • 10
Select™ SPS™ Ultrasonic Scaler Description
Select™ SPS™ Ultrasonic Scaler Description
Select™ SPS™ Ultrasonic Scaler Description cont'd.
®
30K Ultrasonic Inserts are easily interchangeable for various procedures and applications. See enclosed
30K
30K
TM
TM
O-Ring
Provides seal for handpiece
Provides seal for handpiece
coolant.
Finger Grip
Insert Marking
Manufacturer, Date (YYMM),
Frequency, Type, Tip Lot
Foot Control
depressed half
way...
(1st position)
The ultrasonic insert tip is activated
and irrigating water flows.
cont'd.
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the
handpiece into mechanical oscilla-
tions used to activate the insert tip.
IPX1
Foot Control fully depressed...
(2nd position)
This activates the "Boost" Mode.
Tabla de contenido
loading

Este manual también es adecuado para:

SelectSpsCavitron select sps

Tabla de contenido