III. CLEANING - MAINTENANCE
III. 1. Cleaning
Only the outside of the NewLife
Intensity is to be cleaned, with a dry cloth or, if necessary,
®
a damp sponge and clean or soapy water, then thoroughly dried with wipes and an alcohol-based
solution.   A cetone,  solvents  or  any  other  flammable  products  must not be used. Do not use
abrasive powders.
The  removable  dust  filter  (9)  must  be  cleaned  in  soapy  water  after  
approximately 100 hours of use, or when required. Rinse and
wipe  dry.  Fit  a  dry  filter.
III. 2. Everyday disinfection
Due  to  the  presence  of  the  product  filter  inside  the  device,  
everyday disinfection only concerns the external oxygen therapy
accessories:  humidifier,  probes,  nasal  cannulas  (refer  to  the  respective  instructions  for  use).
The use of alcohol-based solutions means that the device must be
switched off.
a - The following minimum guidelines must be respected:
•  Humidifier  :
Daily:
-­  empty  the  water  from  the  humidifier.
-­  rinse  the  humidifier  flask  under  running  water.
-­  Fill  the  humidifier  up  to  the  mark  with  slightly  mineralised  water.
Regularly:
- Disinfect the equipment by immersing them in a disinfectant solution (in general, we recommend
using water containing a small amount of chlorine bleach).
- Rinse and dry.
-­  Check  that  the  humidifier  lid  seal  is  in  good  condition.
• Oxygen administration devices:
Follow the manufacturer's instructions.
b - For each new patient:
The  humidifier  must  be  sterilised  if  possible  or  changed.  The  NewLife
Intensity must be
®
cleaned  and  disinfected  as  per  the  above  instructions.  The  dust  filter  should  preferably  be  changed.  
The entire oxygen administration circuit (oxygen therapy nasal cannulas, etc.) must be changed.
AirSep
Corporation
®
4-GB
loading