Medela Symphony Manual De Instrucciones página 21

Ocultar thumbs Ver también para Symphony:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
5.4 Disposable one-day pump set – Cleaning
1
Check whether there is milk in the
tubing. If so, replace the pump set. If
not, disassemble the pump set into its
individual parts (breast shield,
connector, membrane, membrane cap,
protective membrane and
tubing). You must also remove the
white valve from the connector.
➙ Note that the membrane cap, pro-
tective membrane and tubing must
not be cleaned.
5.5 Disposable one-day pump set
– Disinfection and sterilisation
Do not disinfect or sterilise the
disposable one-day pump set.
2
Rinse the breast shield, the connector and
the white valve with cold, clear drinking-
quality water (approx. 20 °C) to remove
proteins. Clean these parts with plenty
of warm, soapy water (approx. 30 °C) to
wash off fat. Use a commercially available
washing-up liquid, preferably without artificial
fragrances and colouring (5–15% anionic
tenside, nonionic tenside, pH neutral). Rinse
the parts with cold, clear drinking-quality
water (approx. 20 °C).
5.6 Disposable one-day pump set –
Disposal
Dispose of the disposable one-day
pump set after a maximum of eight
pumping sessions within 24 hours.
21
Tabla de contenido
loading

Tabla de contenido