Status: Approved Doc#: AW10001726 Effective Date: nulldate

Section I - Functional Description

This manual describes the operation of the GENIUS 2 Checker/Calibrator. This device has been developed to check the accuracy
of the GENIUS 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the GENIUS 2
Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB flash drive. The
GENIUS 2 Checker/Calibrator will work with all GENIUS 2 Thermometer software revisions.
The GENIUS 2 Checker/Calibrator contains two independently controlled infrared calibration targets that are similar to factory
calibration targets. These "blackbody" targets are designed to have efficient radiative heat transfer and produce infrared radiation
that approximates the human ear at temperatures of 32.2° C (90.0° F ) and 40.6° C (105.0° F ).
11
Figure 1 - Front Panel Layout
Legend
1 Power Connector
2 Power Switch
3
Fuse Holder
4
Low Temperature Target LED
5
Low Temperature Target
6
GENIUS 2 Thermometer Connector
GENIUS ™
2
Checker / Calibrator
1
2
3
F 3.15A 250V
GENIUS
Checker / Calibrator
10
9
7
8
9
10
11
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
4
F/C˚
F/C˚
8
High Temperature Target
High Temperature Target LED
LCD Display
Encoder Knob
USB port
1
5
6
7
Operation and Service Manual