S Series Ultrasound System
User Guide Supplement P13895-06
Introduction .............................................................................................................................1
Imaging .....................................................................................................................................1
Safety .........................................................................................................................................4
Introduction
This user guide supplement does the following:
• Updates acoustic output information for the S Series™ ultrasound system
• Describes SonoMBe
transducers:
• C60x/5-2 MHz
• HFL38x/13-6 MHz
• HFL50x/15-6 MHz
• L25x/13-6 MHz
• L38xi/10-5 MHz
Imaging
2D imaging
WARNING:
To avoid incorrect needle placement when MBe is on:
• Using movement and fluid injection, verify the needle-tip location and trajectory.
• Note that linear structures are enhanced only in an outlined portion of the image.
• Note that the beam divergence of a curved array transducer may prevent a
™ multi-beam imaging, enhanced (MBe), available on the following
MBe enhances linear structures within a selected angle range on the ultrasound
plane. Linear structures outside the selected angle range or the ultrasound
— such as a bent needle — may be less apparent.
plane
The area outside the outline remains unchanged. (See
segment of the needle shaft from showing in the image. (See
tip may not show.
Figure
2.)
Figure
3.) The needle
1