Metrohm 765 Dosimat Instrucciones Para El Uso página 44

Idiomas disponibles

Idiomas disponibles

4.2. Control commands
If a value is entered which is too high or too low, resp., the Dosimat corrects the value
automatically to rate
4.2.7). To set rate
4.2.26 VUA
Setting rate control to 'up analogue' (control via potentiometer (4)).
A set value (VUP VALUE) is overwritten.
4.2.27 VDA
Setting rate control to 'down analogue' (control via potentiometer (4)).
A set value (VDW VALUE) is overwritten.
4.2.28 VDS VALUE
Value = 0.001 ... 999.999
Setting the dispensing volume in ml.
In modes DIS R and DIS C only, in other modes the command is not accepted and bit 0
of information byte 2 is set to 1.
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
4.2.7).
4.2.29 VPI VALUE
Value = 0.001 ... 49.500
Setting the pipetting volume in ml.
In modes PIP and DIL only, in other modes the command is not accepted and bit 0 of
information byte 2 is set to 1.
40
Cylinder
ml
1
5
10
20
50
or rate
, resp. and sets bit 1 of information byte 2 to 1 (see
max
min
it is therefore possible to always enter 150 ml/min.
max ,
Rate up analog, live
Rate down analog, live
Dispensing volume, not live
Cylinder
ml
1
5
10
20
50
or volume
max
min,
Pipetting volume, not live
rate
rate
min
ml/min
ml/min
0.001
0.005
0.010
0.020
0.050
150
Volume
Volume
min
ml/min
ml/min
0.001
999.999
0.001
999.999
0.001
999.999
0.002
999.999
0.005
999.999
resp. and sets bit 1 of information byte 2 to 1 (see
max
3
15
30
60
max
765 Dosimat
loading