Fully insert the phase contrast condenser into the holder, pushing it until it is well inserted:
Rotate the condenser annuli disc to the "0" position (brightfield).
Completely open the aperture diaphragm:
Turn on the light and put a sample slide on the stage (an opaque specimen will make alignment easier).
Using the 10x objective, center your sample (moving the stage) and focus using the coarse and fine focus
knobs.
Rotate the condenser height adjustment knob until you obtain a uniform white illumination on the sample.
Important: the optimal position corresponds to the top lens of the condenser 1-2mm below the bottom side
of the slide.
Rotate the condenser annuli disc to the "10" position.
Page 6
Completely open this diaphragm