Safety Symbols - koban KMD 03 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

KMD 03
Digital multimeter
tip (metal part).
• When measuring, if the voltage to be measured is more than 60 V DC or
30 V AC (RMS), always keep your fingers always behind finger protection
device.
• Don't measure voltage greater then 600 V.
• For manual measuring range, when the value to be measured is unknown,
select, the highest measuring range.
• Before rotating conversion switch to change measuring function, remove
probe from the circuit to be measured.
• Don't measure resistors, capacitors, diodes and circuit connections with
power.
• During tests of current, resistors, capacitors, diodes and circuit connections,
avoid connecting the meter with voltage sources.
• Do not measure capacitance before capacitor is discharged completely.
• Do not use the meter in explosive gas, vapor or dusty environments.
• If you find any abnormal phenomena or failure on the meter, stop using it
immediately.
• Do not use the meter unless the bottom case and the battery cover are
completely fastened in their original places.
• Don't store or use the meter in direct sunlight or high temperature and
high humidity conditions.

Safety Symbols

Note (Important safety information. Refer to the operation
manual)
Can be used for dangerous electric conductor.
Double insulation protection (class II)
CAT IV
According to pulse voltage tolerance protection level
provided by IEC 61010-1 standard overvoltage (installation)
level III and pollution degree 2.
The meter complies with EU standard
Grounding