Change Sensor Type
Prerequisite: The M400 transmitter and another sensor type are mounted and electrically con-
nected.
If you change the sensor type e.g. a Conductivity sensor with a pH sensor, perform the follow-
ing procedure:
1. Built up connection.
2. Select Sensor Setup.
Menu path: Detailed Setup > Measurements > Channel Setup > Sensor Setup.
3. Start "Sensor Setup".
4. Select the new sensor type for the Sensor Type parameter.
5. Load configuration from device.
6. The configuration in the configuration tool or in the HART handheld is updated. If you
have selected the correct sensor type, the "Verify" menu will be shown.