User Controls

ON / OFF Button
Press once to turn "ON"; Press and hold for one second to turn "OFF".
Audible Alert Button
Pressing this button will toggle the OxyGo FIT® 's breath detection audible
alert on and off.
Breath Detection Alert Mode. The OxyGo FIT® will alert with
audible and visual signals for "no breath detected" when this mode
is enabled and no breath has been detected for 60 seconds.
At 60 seconds, the device will enter into auto pulse mode and once
another breath is detected, the device will exit auto pulse mode and
deliver normally on inspiration. The display's mode indication area will
show a bell icon, flashing yellow light and display message when the
alert is enabled.
If power is lost, the breath detection audible alert remains set in the
user preferred mode.
Flow Setting Control Buttons
Use the – or + flow setting control buttons to select the setting
as shown on the display. There are three settings, from 1 to 3.

User Interfaces

Display
This screen displays information regarding flow setting, power status,
battery life and errors. If you would like to change the language on the
LCD Screen contact your equipment provider for instructions. Actual
display appearance may vary.