4. Install a replacement fuse, OxyGo FIT® RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Cigarette Adapter Plug
CAUTION For continued protection against risk of fire, only use fuse specified.
OxyGo FIT® Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
WARNING Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or
C
!
impair equipment performance and will void your warranty.
1. Turn off the OxyGo FIT® concentrator by pressing the power button to shut down the
device.
2. Remove the OxyGo FIT® concentrator from the carry case, if applicable..
3. Remove the battery from the OxyGo FIT®
concentrator.
4. Place the OxyGo FIT® concentrator on its side so that
the underside is visible. The metal column assembly
can be seen on one side of the device.
Fuse
Retainer Tip
Cleaning, Care and Maintenance Chapter 6
Retainer
35